FREE BOOKS

Author's List




PREV.   NEXT  
|<   168   169   170   171   172   173   174   175   176   177   178   179   180   181   182   183   184   185   186   187   188   189   190   191   192  
193   194   195   196   197   >>  
unknown electromotive force of the couple. The two are connected with opposite polarity, or so that the two e.m.f.s oppose one another. So long as one is stronger than the other a current will flow through the couple; when the two are equal no current will flow. Figure 107 shows the wiring of the potentiometer in its simplest form. The thermo-couple is at _H_, with its polarity as shown by the symbols + and -. It is connected with the main circuit of the potentiometer at the fixed point _D_ and the point _G_. [Illustration: FIG. 110.--Simple potentiometer.] A current from the dry cell _Ba_ is constantly flowing through the main, or so-called potentiometer circuit, _ABCDGEF_. The section _DGE_ of this circuit is a slide wire, uniform in resistance throughout its length. The scale is fixed on this slide wire. The current from the cell _Ba_ as it flows through _DGE_, undergoes a fall in potential, setting up a difference in voltage, that is, an electromotive force, between _D_ and _E_. There will also be electromotive force between _D_ and all other points on the slide wire. The polarity of this is in opposition to the polarity of the thermo-couple which connects into the potentiometer at _D_ and at _G_. By moving _G_ along the slide wire a point is found where the voltage between _D_ and _G_ in the slide wire is just equal to the voltage between _D_ and _G_ generated by the thermo-couple. A galvanometer in the thermo-couple circuit indicates when the balance point is reached, since at this point the galvanometer needle will stand motionless when its circuit is opened and closed. [Illustration: FIG. 111.--Standard cell potentiometer.] The voltage in the slide wire will vary with the current flowing through it from the cell _Ba_ and a means of standardizing this is provided. _SC_, Fig. 111, is a cadmium cell whose voltage is constant. It is connected at two points _C_ and _D_ to the potentiometer circuit whenever the potentiometer current is to be standardized. At this time the galvanometer is thrown in series with _SC_. The variable rheostat _R_ is then adjusted until the current flowing is such that as it flows through the standard resistance _CD_, the fall in potential between _C_ and _D_ is just equal to the voltage of the standard cell _SC_. At this time the galvanometer will indicate a balance in the same way as when it was used with a thermo-couple. By this operation the current in the slide wire _DGE_ h
PREV.   NEXT  
|<   168   169   170   171   172   173   174   175   176   177   178   179   180   181   182   183   184   185   186   187   188   189   190   191   192  
193   194   195   196   197   >>  



Top keywords:
potentiometer
 

current

 

couple

 

voltage

 

circuit

 

thermo

 
polarity
 

galvanometer

 

connected

 

electromotive


flowing
 

standard

 
Illustration
 
balance
 

points

 

potential

 
resistance
 

standardizing

 

provided

 

cadmium


opened

 

reached

 

oppose

 

needle

 

closed

 
motionless
 

Standard

 

standardized

 

operation

 

adjusted


generated

 

thrown

 
series
 
rheostat
 
variable
 

constant

 

moving

 

uniform

 

unknown

 
simplest

wiring

 

length

 

section

 

ABCDGEF

 
symbols
 

Simple

 

called

 

constantly

 
Figure
 

undergoes