ince only with solutions that are entirely clear and bright can accurate polarimetric observations be made. When a sufficient quantity of the clear liquid has passed through the filter, the 200 mm. observation tube is filled with it. The 100 mm. tube should never be used except in rare cases, when notwithstanding all the means used to effect the proper decolorization of the solution, it is still too dark to polarize in the 200 mm. tube. In such cases the shorter tube may be used, and its reading multiplied by two. The zero deviation must then be determined and applied to the product. This will give the reading which would have been obtained if a 200 mm. tube could have been used, and it only remains to apply the correction determined by the use of the control plate as previously described. Example: Solution reads in 100 mm. tube 47.0 Multiplied by 2 2.0 ---- Product 94.0 Zero reads plus 0.3 0.3 ---- Solution would read in 200 mm. tube 93.7 Reading of control plate 90.4 Sugar value of control plate 90.5 ---- Instrument too low by 0.1 Add 0.1 to 93.7 ---- Correct polarization of solution 93.8 Before filling the tube it must either be thoroughly dried by pushing a plug of filter paper through it, or it must be rinsed several times with the solution itself. The cover glasses must also be clean and dry, and without serious defects or scratches. Unnecessary warming of the tube by the hand during filling should be avoided; it is closed at one end with the screw cap and cover glass, and grasped by the other end with the thumb and finger. The solution is poured into it until its curved surface projects slightly above the opening, the air bubbles allowed time to rise, and the cover glass pushed horizontally over the end of the tube in such a manner that the excess of liquid is carried over the side, leaving the cover glass exactly closing the tube with no air bubbles beneath it, and with no portion of the liquid upon its upper surface. If this result is not attained, the operation must be repeated, the cover glass being rubbed clean and dry, and the solution ag
PREV.   NEXT  
|<   83   84   85   86   87   88   89   90   91   92   93   94   95   96   97   98   99   100   101   102   103   104   105   106   107  
108   109   110   111   112   113   >>  



Top keywords:

solution

 

liquid

 

control

 
bubbles
 
surface
 

filling

 

determined

 

reading

 

Solution

 

filter


avoided

 

closed

 

rinsed

 
pushing
 
scratches
 

defects

 
Unnecessary
 

warming

 

glasses

 
slightly

beneath

 

portion

 

closing

 

carried

 

leaving

 

rubbed

 
repeated
 

operation

 

result

 
attained

excess

 

manner

 
poured
 

finger

 
grasped
 

curved

 

projects

 

pushed

 

horizontally

 

allowed


opening

 

Multiplied

 

effect

 

proper

 

decolorization

 
notwithstanding
 
multiplied
 

shorter

 

polarize

 
accurate